DD Test
Default test time of 30 minutes, followed by a 1 minute discharge time.
Terminating the test prematurely still allows the test results to be
calculated and displayed.
Timer
Digital Indicator
T3
Test Duration
Blank
T3 time
DD value
T3 time
R at T3 time
T3 time
I at T3 +1 minute
PI (T1 / T2)
T2 time
µF
T3 time
Note: A capacitance value reading is only available if the test has run for
more than one minute, and provided the reading was not over-range.
Voltage Indicator
Other
Final test voltage
Voltage at T3 time
DD cursor
flashes
Voltage at T3 time
Voltage at T3 time
PI cursor
Voltage at T2 time
flashes
Voltage at T3 time
SV Test
Test time set at 5 minutes only. Terminating the test prematurely may
cause some results to be displayed as - - -.
Digital Indicator
Timer
Test Duration
R (at 1 minute)
1,00 minute
R (at 2 minutes)
2,00 minute
3,00 minute
R (at 3 minutes)
4,00 minute
R (at 4 minutes)
R (at 5 minutes)
5,00 minute
Test Duration
Note: A capacitance value reading is only available if the test has run for
more than one minute, and provided the reading was not over-range
Voltage Indicator
Blank
Final test voltage
Voltage at 1 minute
Voltage at 2 mins
Voltage at 3 mins
Voltage at 4 mins
Voltage at 5 mins
µF
Final test voltage