Tip ejector assembly (Fig. 6A)
1. Remove the tip ejector button.
2. Use the screwdriver to turn the metal arbor right to
screw the ejector in and adjust its length.
Tip ejector adjustment (Fig. 6A)
1. Remove the ejector pushbutton.
2. Use the screwdriver to increase or decrease the ejec-
tor length.
After ejector assembly or adjustment, put the ejector but-
ton back on its place.
If above described method of ejector adjustment is not
sufficient or the diameter of the ejector opening is too
large to dispose the tip it is necessary to put the ejector
cap "M" onto the ejector, (Fig. 6B).
- in 2 and 10 μl pipettes (Fig. 6C)
The 2 μl - 10 μl pipettes may require the use of an ejector
cap "M" to efficiently eject certain brands of pipette tips.
Simply place the cap on the bottom of the pipette shaft
and slide the cap upwards until it surrounds the bottom
of the tip ejector.
The pipette is a high quality instrument which offers
excellent accuracy and precision.
The accuracy and precision of liquid volume depend on
the quality of tips used. The values for accuracy and
precision given in the table below were obtained using
HTL tips. Only those tips guarantee correct operation of
the pipettes and ensure accuracy and precision of liquid
sampling.
Volume
Model
Cat. no.
[μl]
0.2
LMP2
5661
1.0
Max
2.0
Min
0.5
LMP10
5662
5.0
Max 10.0
Min
LMP20
5663
10
Max
20
Min
LMP50
5667
25
Max
50
Min
10
LMP100
5664
50
Max
100
3
Accuracy
Precision
Tip
[%]
[%]
μl
± 12.0
± 6.0
± 2.7
± 1.3
± 1.5
± 0.7
10 μl
± 4.0
± 2.8
± 1.0
± 0.6
± 0.5
± 0.4
2
± 3.0
± 1.5
± 1.0
± 0.5
± 0.8
± 0.3
5
± 2.5
± 2.0
200 μl
± 1.0
± 0.6
± 0.8
± 0.4
± 1.6
± 0.80
± 0.8
± 0.24
± 0.8
± 0.20
Min
20
LMP200
5665
100
Max
200
Min
50
LMP250
5660
125
Max
250
Min
100
LMP1000
5666
500
Max 1000
Min
500
LMP5000
5668
2500
Max 5000
Min 1000
LMP10000
5669
5000
Max 10000
The accuracy and precision are obtained with HTL tips,
using a gravimetric method, performing at least 10 mea-
surements of distilled water at a temperature of 20±1°C,
according to EN ISO 8655 standard.
The pipette design enables the user to perform the
recali bration process according to the rules presented in
section 8.

2 - SETTING THE VOLUME

The counter displays three figures to be read from top to
bottom. Additional to the figures on the lower wheel are
printed graduations to enable a volume setting in the
range increment of each pipette model.
Examples of the meaning of the black and red figures:
Pipettes LMP2
Red figures at the bottom = 1/100 μl
Increment = 0.002 μl
Pipettes LMP10, LMP20,
Red figures at the bottom = 1/10 μl
Increment = 0.02 μl
Pipettes LMP50, LMP100,
LMP200, LMP250
Black figures only = 1 μl
Increment = 0.2 μl
ENGLISH
± 1.2
± 0.60
200 μl
± 0.8
± 0.25
± 0.6
± 0.20
± 1.0
± 0.4
300 μl
± 0.8
± 0.3
± 0.6
± 0.3
± 1.6
± 0.40
± 0.7
± 0.20
1000 μl
± 0.6
± 0.15
± 1.2
± 0.50
5000 μl
± 0.6
± 0.20
± 0.5
± 0.15
± 2.5
± 0.6
10000 μl
± 0.8
± 0.3
± 0.5
± 0.2