Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 30
• Liquid should never enter the pipette shaft. To prevent
this:
- Press and release the pushbutton slowly and smoothly.
- Never turn the pipette upside down.
- Never lay the pipette on its side when there is liquid in
the tip.
• Never force the volume setting beyond its recommended
limits.
• When pipetting liquids with temperatures different from
the ambient temperature, it is reco mmended to pre-
rinse the tip several times before use.
• Do not pipette liquids with temperatures above 70°C.
• When pipetting acids or corrosive solutions which emit
vapours, it is recommended to disassemble the shaft
and to rinse the piston and seal with distilled water after
finishing the pipetting operation.

8 - RECALIBRATION

HTL pipettes are calibrated by gravimetric me thod, using
HTL tips and distilled water, at the temperature 20±1°C,
according to EN ISO 8655 standard.
If during pipette operation you find that the accuracy
error (the difference between the real aspirated volume
and the preset volume) exceeds the permissible value
given in the table in section 1, the pipette recalibration
procedure should be carried out.
Before starting the recalibration it is necessary to check
whether the following requirements have been fulfilled
during error determination:
• the ambient temperature, and the temperature of the
pipette, tips and water was identical
• the density of the liquid used is close to that of distilled
water
• the balance with appropriate sensitivity has been used.
Volume checked [μl]
0.1 - 10
10 - 100
> 100
• mg/μl conversion factor has been taken into ac count
• the requirements given in sections 3 and 7 have been
fulfilled
9
Balance sensitivity [mg]
≤ 0.001
≤ 0.01
≤ 0.1
If the above conditions are satisfied and the ac cu racy error
for selected volume given in section 1 exceeds the per-
missible value, the pipette recalibration procedure
should be carried out.
The recalibration can be performed within one full turn
of the key to the right or to the left only.
Recalibration conditions:
• Ambient temperature and the temperature of the
pipette, tips and liquid should be within the range 20-
25°C and stabilised during weighing within ±0.5°C
• Measurements should be conducted using distilled
water
• Balance sensitivity should be suitable for the vo lume to
be controlled
Recalibration procedure:
• Set the dose volume depending on the pipette volume
according to the following table:
Range
Preset
Model
of the pipette
volume
volumes [μl]
[μl]
LMP2
0.1 - 2
0.2
LMP10
0.5 - 10
0.5
LMP20
2 - 20
2
LMP50
5 - 50
5
LMP100
10 - 100
10
LMP200
20 - 200
20
LMP250
50 - 250
50
LMP1000
100 - 1000
100
LMP5000
500 - 5000
500
LMP10000
1000 - 10000
1000
• Perform 5 aspirations, weigh each one and calculate
the average value of the aspirations.
• Calculate average aspirated volume in [μl] multiplying
the average aspiration amount [mg] by the distilled
water density coefficient [μl/mg], which depends on
temperature and pressure according to the following
table:
ENGLISH
Volume change
ΔV for full turn
Permissible
of the
volumes
calibration key
[μl]
[μl]
(24 increments)
0.176 - 0.224
0.06
0.48 - 0.52
0.33
1.94 - 2.06
0.63
4.875 - 5.125
2.50
9.84 - 10.16
2.50
19.76 - 20.24
6.30
49.5 - 50.5
6.30
98.4 - 101.6
25.00
494 - 506
125.00
975 - 1025
250.00
10

Publicidad

Tabla de contenido
loading

Tabla de contenido