7 - REPLACING THE FILTER
CAUTION:
The work safety instructions given in section 2 shall be observed when disassembling the
pipette controller.
The filter replacement is necessary, if drawing efficiency deterioration is observed.
The direct reason may be dirty filter after a long period of use.
In order to replace the filter:
• Remove the pipette.
• Unscrew the nosepiece (fig. 4A).
• Remove the membrane filter (fig. 4A) and the pipette holder (fig. 4B).
• Rinse the holder using a wash bottle (fig. 4C).
• Blow liquid out of the holder and leave it until it is completely dry.
• Install new membrane filter (fig. 4D) and assemble the device in reverse order.
8 - CHARGING THE BATTERIES
CAUTION!
The SWIFTPET PRO may be charged only with the original charger. The mains voltage must
conform with the specification on the charger.
Using other chargers than the original one may damage the battery.
The pipette controller is delivered with 3 NiMH type AAA batteries with a capacity of 900 mAh.
The batteries can be replaced easily if necessary after the battery cover is taken off (fig. 1K).
The manner in which the batteries are arranged is shown in fig. 8.
Charging:
1. Charging temperature: 10 °C to 35 °C.
2. Charging the battery is carried out through a charger (power supply) by direct connection
to the mains, or indirectly by means of a charging stand (fig. 1M). Batteries charging is indi-
cated by successive lighting of "bars".
3. Full charging time: 7-8 hours.
4. The batteries are charged when all 3 "bars" are displayed simultaneously (fig. 5.1).
When the batteries are charged, the charging circuit disconnects automatically.
The service life of the batteries: approx. 1000 charging cycles, if used correctly.
It is not possible to overcharge the batteries if all instructions of the manufacturer are followed.
ENGLISH
6