Calibrating the Toxic Gas Instruments
(CO and H 2 S)
To enter the calibration mode, make sure you are in fresh,
uncontaminated air. See FIGURE 4-1 for more details.
1. In normal operating mode, press the TEST button.
• "TEST" "GAS" "?" displays.
2. Press and hold the TEST button for three seconds when the
"TEST" "GAS" "?" displays.
• "TEST" "CAL" screen displays
• After three seconds, "FAS" "?" displays asking the user if a
fresh air setup/calibration is desired.
3. Press the TEST button to enter zero calibration.
NOTE: Otherwise, the instrument will return to the normal
operating mode.
4. During zero calibration:
• the hour glass and "FAS" display
• If the instrument successfully calibrates:
"OK" displays
• If the instrument does not successfully calibrate:
"ERR" displays
instrument returns to normal operating mode after five
seconds.
5. Once the instrument successfully calibrates and the "OK" displays,
press the TEST button to enter calibration.
"CAL" "?" displays.
6. While the "CAL" "?" displays, press the TEST button to enter the
Gas Calibration mode.
The current expected test gas is shown (in ppm).
4-2