Chapter 1,

Instrument Safety and Certifications

The Altair Single Gas Detector is:
For use by trained, qualified personnel
For use as a warning device only and not for measuring specific
gas concentrations
To be used when performing a hazard assessment to:
Assess potential worker exposure to Specific toxic gases
for which a sensor is installed
Oxygen-deficient or oxygen-rich atmospheres
NOTE: Although the Altair unit will detect up to 25% oxygen, all
Altair versions are not certified or classified for use in
atmospheres containing more than 21.0% oxygen
Read and follow all instructions carefully.
Do not use this detector to sample the instrument's
specified gas in gases other than air.
Perform an alarm function check and a bump test before
each day's use. If the instrument fails either check, the
instrument must be taken out of service.
Recheck response and alarm function if the instrument is
subjected to physical shock.
The instrument is factory-sealed and contains no user
serviceable parts. Substitution of components may impair
intrinsic safety.
This unit contains a lithium battery; dispose of according to
local regulations.
Do not rely on the vibrating alarm in cold temperatures
(<0°C) as the vibrating alarm may cease to operate under
these conditions.
Use only to detect a gas for which a sensor is installed.
Do not block sensor.
Do not use pressurized air to clean the sensor holes.
All instrument readings and information must be interpreted
by someone trained and qualified in interpreting instrument
readings in relation to the specific environment, industrial
practice and exposure limitations.

" WARNING

1-1