Chapter 3,
Altair Function Checks
Confidence and Heartbeat Indicator
•
The alarm lights and heartbeat indicator on the display will
flash approximately every 60 seconds to indicate that the
Altair Detector is operating.
Alarm Test
•
Check before each day's use.
•
Press the TEST button momentarily. A one-second test of the
alarms will occur; this includes the:
•
display
•
alarm lights
•
vibrator
•
horn.
•
If these items do not activate, remove the instrument from service.
Bump Test
•
Check before each day's use.
•
Press the TEST button momentarily:
•
oxygen versions will display the current oxygen reading;
calibrate the unit if it reads other than 20.8%.
•
"TEST" "GAS" "?" will display.
•
alarm test will activate.
•
Press the TEST button again while "TEST" "GAS" "?" is displayed
to activate the Bump Test mode.
•
the hourglass and "GAS" will display.
•
Apply gas only AFTER the hourglass and "GAS" are displayed.
•
If gas is detected, the display will indicate "OK".
NOTE:
See TABLE 3-1 for applied gases.
3-1