1. Instrument Safety and Certifications

The SIRIUS Multigas Detector is for use by trained and qualified personnel. It is designed
to be used when performing a hazard assessment to:
Assess potential worker exposure to combustible and toxic gases and vapours
Determine the appropriate gas and vapor monitoring needed for a workplace.
The SIRIUS Multigas Detector can be equipped to detect:
Combustible gases and certain combustible vapours
Volatile organic compounds (VOCs)
Oxygen-deficient or oxygen-rich atmospheres
Specific toxic gases for which a sensor is installed.
ATTENTION:
Read and follow all instructions carefully.
Check calibration before each day's use and adjust if necessary.
Check calibration more frequently if exposed to silicone, silicates, lead-containing
compounds, hydrogen sulfide, or high contaminant levels.
Recheck calibration if unit is subjected to physical shock.
Use only to detect gases/vapours for which a sensor is installed.
Do not use to detect combustible dusts or mists.
Make sure adequate oxygen is present.
Do not block pump sample inlet and sensors.
Use only Teflon sampling lines for reactive gases such as Cl
for semivolatile organic compounds such as gasoline and jet fuels.
Do not place end of sampling line in liquids.
Use only MSA-approved sampling lines.
Do not use silicone tubing or sampling lines.
Wait for accurate reading; response times vary, based on gas/vapor and length of
sampling line.
Have a trained and qualified person interpret instrument readings.
Account for sensor reproducibility.
Properly identify the VOC gas being measured before using VOC response factors
or setting alarm values (exposures, STEL, TWA).
Recognize that the VOC Auto-range displays readings in increments of 100 ppb.
Ensure installed PID bulb corresponds to the PID bulb setting on the instrument
display.
Do not recharge Lithium Ion battery or change Alkaline batteries in a combustible
atmosphere.
Do not alter or modify instrument.