1. Instrument Safety and Certifications
The SIRIUS Multigas Detector is for use by trained and qualified personnel. It is designed
to be used when performing a hazard assessment to:
•
Assess potential worker exposure to combustible and toxic gases and vapours
•
Determine the appropriate gas and vapor monitoring needed for a workplace.
The SIRIUS Multigas Detector can be equipped to detect:
•
Combustible gases and certain combustible vapours
•
Volatile organic compounds (VOCs)
•
Oxygen-deficient or oxygen-rich atmospheres
•
Specific toxic gases for which a sensor is installed.
ATTENTION:
•
Read and follow all instructions carefully.
•
Check calibration before each day's use and adjust if necessary.
•
Check calibration more frequently if exposed to silicone, silicates, lead-containing
compounds, hydrogen sulfide, or high contaminant levels.
•
Recheck calibration if unit is subjected to physical shock.
•
Use only to detect gases/vapours for which a sensor is installed.
•
Do not use to detect combustible dusts or mists.
•
Make sure adequate oxygen is present.
•
Do not block pump sample inlet and sensors.
•
Use only Teflon sampling lines for reactive gases such as Cl
for semivolatile organic compounds such as gasoline and jet fuels.
•
Do not place end of sampling line in liquids.
•
Use only MSA-approved sampling lines.
•
Do not use silicone tubing or sampling lines.
•
Wait for accurate reading; response times vary, based on gas/vapor and length of
sampling line.
•
Have a trained and qualified person interpret instrument readings.
•
Account for sensor reproducibility.
•
Properly identify the VOC gas being measured before using VOC response factors
or setting alarm values (exposures, STEL, TWA).
•
Recognize that the VOC Auto-range displays readings in increments of 100 ppb.
•
Ensure installed PID bulb corresponds to the PID bulb setting on the instrument
display.
•
Do not recharge Lithium Ion battery or change Alkaline batteries in a combustible
atmosphere.
•
Do not alter or modify instrument.