Rinse the internal canal of the insert
with distilled water injected under pressure
(3,8 bar) for at least 10 seconds to
eliminate all residues;
EN
Place the insert in the ultrasound tank
submerged by the enzymatic detergent
solution at 40°C ±2°C, for at least 10
minutes;
Remove the insert from the ultrasonic tank
and rinse with distilled water;
Delicately brush the surface of the insert
again with the brush with soft nylon
bristles;
Rinse the internal canal of the insert
with distilled water injected under pressure
(3,8 bar) for at least 10 seconds to
eliminate all residues;
Process validated by an independent institute with
enzymatic detergent Enzymec (Mectron).