ONLY AUTOMATED CYCLE

NOTE: Place the accessories in the washing
machine so that dead zones do not arise
and the water can properly drain. Also,
make sure that the reusable inserts are
properly held in place in the washing basket
and cannot move during the washing
process, as shocks could damage them.
WARNING: Avoid overloading. Failure
to comply overloading compromises
cleaning effectiveness.
CAUTION: Do not kink the handpiece
cord; this may damage it.
_ _
Procedure validated with Miele PG8535 washer/
disinfector.
Programm Miele DES-VAR-TD.
Alcaline detergent: neodisher® FA (0.2 % v/v).
Neutralizer liquid: neodisher® Z (0.1 % v/v)
06
CLEANING OF THE ACCESSORIES
1
2
+
-
WARNING: Once the cleaning cycle
in thermodisinfector has been completed,
the handpiece keeps the washing tempera-
ture for a long time. When removing the
handpiece from the thermodisinfector take
proper measures to avoid damages for the
user.
CAUTION: Due to its shape, the hand-
piece can roll. For this reason, when it is
not in use, it shall always be placed on its
support.
Lay the accessories in a metallic tray.
Connect the suitable adapter (supplied as
optional) to the handpiece cord connector
and then to the cleaning water jet
connections of the washer/disinfector.
Repeat the same step for the inserts and
for the peristaltic pump tubes, connecting
them to the suitable adapters. (suuplied as
optional).
Sequence and parameters applicable to
the cycle:
CLEANING
- 1 min, Pre-washing with cold water;
- 5 min, Washing with alcaline detergent at
55°C ±2°C;
- 1 min, Neutralization with proper solution
(1/3 cold water, 2/3 warm water);
- 1 min, Rinsing with water (1/3 cold water,
2/3 warm water);
DISINFECTION
- 5 min, Thermodisinfection at 93°C with
demineralized water.
The automated thermodisinfection was
not experimentally tested. According to
standard ISO 15883-1, Table B.1 [4] the
thermodisinfection at a temperature of
90°C for 5min determines an A0 value of
3000.