06.1.2
CAUTION: The following automated
cleaning cycle and the thermodisinfection
must always be preceded by the pre-
cleaning steps with enzymatic detergent.
EN
NOTE: Place the accessories in the
washing machine so that dead zones do
not arise and the water can properly drain.
Also, make sure that the reusable inserts
are properly held in place in the washing
basket and cannot move during the
washing process, as shocks could damage
them.
WARNING: Avoid overloading. Failure
to comply overloading compromises
cleaning effectiveness.
Lay the accessories in a metallic tray and
place it in thermodisinfector;
Sequence and parameters applicable to
the cycle:
CLEANING
- 1 min, Pre-washing with cold water;
- 5 min, Washing with alcaline detergent at
55°C ±2°C;
- 1 min, Neutralization with proper solution
(1/3 cold water, 2/3 warm water);
- 1 min, Rinsing with water (1/3 cold water,
2/3 warm water);
DISINFECTION
- 5 min, Thermodisinfection at 93°C with
demineralized water.
The automated thermodisinfection was
not experimentally tested. According to
standard ISO 15883-1, Table B.1 [4] the
thermodisinfection at a temperature of
90°C for 5min determines an A0 value of
3000.
20
THERMODISINFECTION
CAUTION: Do not kink the handpiece
cord; this may damage it.
WARNING: Once the cleaning cycle
in thermodisinfector has been completed,
the handpiece keeps the washing tem-
perature for a long time. When removing
the handpiece from the thermodisinfector
take proper measures to avoid damages
for the user.
CAUTION: Due to its shape, the
handpiece can roll. For this reason, when
it is not in use, it shall always be placed on
its support.
1
Procedure validated with Miele PG8535 washer/
disinfector.
Programm Miele DES-VAR-TD.
Alcaline detergent: neodisher® FA (0.2 % v/v).
Neutralizer liquid: neodisher® Z (0.1 % v/v)