Mectron Piezosurgery Manual De Limpieza Y Esterilización página 24

Ocultar thumbs Ver también para Piezosurgery:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

08
DRYING AND LUBRICATION
2
CAUTION: Before starting the
sterilization cycle, make sure that the
insert is thoroughly dry, both externally
and internally. For this purpose, blow
compressed air both externally and
through the internal thru hole; this will
prevent the onset of stains and haloes
on the surface or rust inside the insert.
Thouroughly dry the accessories on all
3
their surfaces, internal and external, using a
or by blowing compressed air, depending
on the case.
Dry the torque wrench using a clean and
4
Lubricate with medical degree lubricant in
5
the indicated point;
CAUTION: do not use lubricants
based on oil or silicon.
EN
23

Publicidad

Tabla de contenido
loading

Tabla de contenido