Automatic Calibration - Astralpool OPTIMA PLUS Manual De Instalación

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 27
Potential Redox Calibration Menu (O.R.P.)
Pressing the CAL key for 3 seconds takes you into the calibration menu. If calibration was excluded during
programming, the following appears on the display:
Calibration
Off
If calibration is active:
Cal
Calibration
Automatic
It is possible to select automatic or manual mode.
-

Automatic calibration:

The buffer solution value appears on the display. Insert the probe in the bottle and press the
countdown necessary to complete calibration will appear on the display. If the alignment quality is below 50%, an
error message appears on the display and you should press
after 4 seconds). If the quality is above 50%, the value is shown on the display and you should press the
to complete the procedure.
Manual calibration:
-
The buffer solution value appears on the display. Insert the probe in the bottle and press the
465 mV should now flash on the display. Insert the probe in your solution and use the
the value of the solution in your possession, then confirm by pressing the
procedure as before
EM00136408
Calibration
Automatic Cal.
Automatic
Buffer Sol. 465mV
-
Calibration
Automatic Cal.
Manual
Buffer Sol. 465mV
rev. 1.0
Wait....
60s
Buffer Sol. 465mV
Quality
100%
Automatic Cal.
||||||
ok 465mV
Error
-
Wait....
+
Buffer Sol. 600mV
Quality
100%
Automatic Cal.
||||||
ok
7 pH
Error
to exit calibration (the pump exits automatically
key and begin the calibration
465mV
60s
7 pH
key. A 60 second
key
key. The value of
keys to display
12