6. TECHNICAL SPECIFICATIONS
This product conforms to the prescriptions of the European directive on low voltage
73/23/EEC (LVD) and to EMC directive 89/336/EEC, amended by 93/68/EEC.
6.1. CHARACTERISTICS
Accuracy is indicated as [% of reading + digit number]. It is referred to the following
reference conditions: 23°C ± 5°C with RH <75%.
DC Voltage
Range
Resolution
400mV
4V
40V
400V
100mV
600V
AC Voltage
Range
Resolution
400mV
4V
40V
400V
100mV
600V
AC Current
Range
Resolution
40A
400A
DC Current
Range
Resolution
40A
400A
Resistance
Range
Resolution
400Ω
0.1Ω
4kΩ
1Ω
40kΩ
10Ω
400kΩ
100Ω
4MΩ
1kΩ
40MΩ
10kΩ
0.1mV
1mV
±(0.8%rdg + 2 dgt)
10mV
±(1.0%rdg + 2 dgt)
1V
±(1.0%rdg + 50 dgt)
0.1mV
1mV
±(1.0%rdg+ 3 dgt)
10mV
±(1.2% rdg + 3 dgt)
1V
0.01A
±(1.5%rdg + 10 dgt) (0.1 – 40A)
0.1A
±(1.5%rdg + 3 dgt) (40.1 – 400A)
0.01A
±(1.5%rdg + 10 dgt) (0.1 – 40A)
0.1A
±(1.5%rdg + 3 dgt) (40.1 – 400A)
Accuracy
±(1.0% rdg + 5 dgt)
±(1.0% rdg + 3 dgt)
±(3.0% rdg + 3 dgt)
Accuracy
Accuracy
(50 ÷ 500Hz)
(at 40Hz - 60Hz)
Accuracy
(50 ÷ 60Hz)
±(2.0%rdg + 10 dgt)
Accuracy
±(2.0%rdg + 10 dgt)
Max. Open Loop
voltage
about 1.5VDC
about 0.45 VDC
EN - 17
ACDC-400
Input
impedance
100MΩ
11MΩ
10MΩ
Input
impedance
100MΩ
11MΩ
10MΩ
Overload
protection
600A rms (60 seconds)
Overload
protection
600A rms (60 seconds)
Overload
protection
600V rms