No additional special skills or training are required to operate the system, but
physicians should be thoroughly familiar with the DyeVert Plus EZ Contrast
Reduction System supporting material including all product labeling. Physicians
may contact Osprey Medical to request training.
DIRECTIONS FOR USE
Note: For Detailed instructions for the Monitor, refer to the Contrast Monitoring
Display Instructions for Use or the Smart Monitor Instructions for Use as
applicable
System Overview
The DyeVert Plus EZ Disposable Kit is for use with the Monitor through wireless
communication. The system allows user input and monitoring of cumulative
volume, injection volume and user inputted threshold. A user selectable toggle
field allows for display of either the % contrast in Smart Syringe, % Contrast
Saved for each injection and % contrast saved cumulative throughout the case.
The system with the DyeVert Plus EZ Module allows contrast modulation to the
patient resulting in a total contrast volume savings.
Threshold Case Screen
Triple Toggle
No Threshold Case Screen
Modulation System
A. Assemble and Prime Manifold
DyeVert
Contrast
Line
Triple Toggle
Note: It is recommended to leave all battery pull tabs in the disposable devices
until all priming is complete and the case is ready to begin. This will ensure
maximum battery life and performance throughout the case.
1) Introduce disposables into sterile field per normal procedure.
2) Attach contrast collection bag to DyeVert Module Reservoir.
3) Attach contrast source line to top of DyeVert reservoir.
4) Attach DyeVert contrast line to contrast port on manifold. Ensure swivel luer
connection is tightened and securely attached to manifold.
5) Attach DyeVert T-Connector to manifold.
6) Turn DyeVert 1-Way Stopcock to "OFF" (parallel to manifold) to isolate the
DyeVert Module Reservoir.
7) Remove air from Smart Syringe and attach to DyeVert T-Connector.
8) Prime manifold and all lines per normal procedure.
B. Prime DyeVert Plus EZ Module With Contrast
1) Turn manifold contrast line stopcock to aspiration position.
2) Aspirate approximately 6ml of contrast into Smart Syringe.
3) Turn manifold contrast line stopcock to "OFF" to Smart Syringe.
4) Turn the DyeVert stopcock to "ON" to savings position (parallel to diversion
line).
Attach contrast line
to luer fitting
Attach
Contrast
Collection
Bag
T-Connector
Stopcock
"OFF"
8287-D