SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
NEVER use tools of any kind to connect/disconnect the cylinder
and the compressor. Otherwise, severe injury and/or damage may
occur.
DO NOT drop oxygen cylinders. Use two hands when
handling/transporting oxygen cylinders. Otherwise, injury or
damage may occur.
NOTE: For this procedure, refer to FIGURE 4.2, FIGURE 4.3 on page 16,
FIGURE 4.4 on page 17 and FIGURE 5.2 on page 21.
1.
Examine the cylinder contents gauge. If the cylinder pressure is less
than 10342 kpa (1500 psig), proceed to STEP 2 (FIGURE 4.4).
NOTE: If the cylinder pressure is greater than 10342 kpa (1500 psig), DO NOT
attempt to top off this cylinder. It may not fill.
2.
Set the regulator flow indicator on the cylinder to the "0" position
(FIGURE 5.2).
3.
For models HF2RE9ES only - Slowly open cylinder valve fully
(FIGURE 4.2).
Regulator
Assembly
Cylinder
Valve
4.
Remove the cylinder and connector fillport covers.
5.
Momentarily push DOWN on the outer ring (sleeve) of the connector
fillport to reset the connector as shown in FIGURE 4.3.
Part No 1167426
WARNING
FIGURE 4.2 Opening Cylinder Valve
15
NOTE: Turn valve
counterclockwise
to open.
HomeFill
®