Invacare HomeFill HF2RE9 Manual Del Usuario página 13

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 49
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
External Examination
NOTE: For this procedure, refer to FIGURE 4.1 on page 14.
1.
Examine the outside of the cylinder for the following conditions, and
replace the cylinder if they exist:
Dents or dings
Arc burns
Oil or grease
Any other signs of damage that might cause a cylinder to be
unacceptable or unsafe for use.
2.
Examine the cylinder for evidence of fire or thermal damage. Evidence
includes charring or blistering of the paint, or other protective coating
or heat sensitive indicator. If fire or thermal damage is found, replace
the cylinder.
3.
Inspect the cylinder fillport for the following:
Debris, oil or grease
Noticeable signs of damage
Signs of corrosion inside the valve
Signs of excessive heat or fire damage
Signs of pressure leaks
NOTE: DO NOT use the cylinder if any of these items are found. Contact
your dealer/Home Care equipment provider for replacement.
Part No 1167426
13
HomeFill
®

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Homefill hf2re9es

Tabla de contenido