SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Use this cylinder in accordance with the HomeFill owner's manual.
Open valve slowly, Close valve after each use and when empty.
Secure cylinder during storage and use. No smoking in cylinder
area. Keep away from heat, flame or spark. Keep out of the reach
of children. DO NOT drop.
Filling of cylinder is performed by Invacare HomeFill oxygen
system only.
Cylinders that have been refinished and/or exposed to elevated
temperatures MUST be hydrostatically tested before filling.
Cylinder temperature should not exceed 52° C (125° F).
Aluminum cylinders subject to the action of fire or heated
temperatures in excess of 177° C (350° F) MUST be withdrawn
from service and condemned by trained personnel. DO NOT use
caustic paint strippers. Valve and safety relief devices MUST be
removed and replaced by trained and authorized personnel. DO
NOT alter or change valve assembly.
All cylinders MUST be inspected before attempting to fill;
otherwise, injury or damage may occur.
NEVER smoke in an area where oxygen is being administered.
NEVER use near any type of flame or flammable/explosive
substances, vapours or atmosphere.
The Hydrostatic Testing Date
Aluminum cylinders MUST undergo testing every five years.
DO NOT fill cylinders that have not been tested in the past five
years. Otherwise, serious personal injury may result. Contact your
dealer/Home Care equipment provider for replacement.
HomeFill
®
WARNING
WARNING
12
Part No 1167426