Medtronic SYNCHROMED II 8637 Manual De Implantación página 28

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Filename Date Time
UC200xxxxxx EN
4.6 x 6.0 inches (140 mm x 197 mm)
prescribed fluid
Figure 9. Open clamp and inject into pump reservoir.
#
Caution: Do not prematurely activate the pump reservoir valve. Activation of the
pump reservoir valve seals the pump reservoir valve closed. Unusual resistance
or the inability to inject the entire fill volume may indicate activation of the pump
reservoir valve. If the valve closes, a portion of the reservoir contents must be
delivered or removed before filling can be completed. Procedural delays can occur.
To prevent activation of the pump reservoir valve during emptying and filling
procedures:
completely aspirate all contents of the pump reservoir before filling.
do not allow air into the pump reservoir through an open needle in the septum
or an unclamped extension.
do not exceed the maximum reservoir volume indicated in the pump labeling.
5. If you have activated the reservoir valve, complete steps a through f below. Otherwise,
proceed to step 6.
a. Discontinue injection, close the clamp, and remove the needle from the septum.
b. Remove the syringe with prescribed fluid and attached filter.
c. Attach an empty 20-mL syringe to the extension tubing set (needle attached).
26 English
8637 2009-10
2009-10
Syringe with
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Template version: 03-03-2010
Filter
Open clamp
M221311A028 Rev A

Publicidad

Tabla de contenido
loading

Tabla de contenido