ELECTROMAGNETIC COMPATIBILITY
12.1 GUIDE AND DECLARATION OF THE MANUFACTURER - ELECTROMAGNETIC EMISSIONS
The equipment known as "LED Source" is intended to be used in the electromagnetic environment specified below.
Users should make sure it is used in this environment.
Emission trial
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic current emissions
EN 61000-3-2
Voltage fluctuations / Flicker
EN 61000-3-3
12.2 GUIDE AND DECLARATION OF THE MANUFACTURER - ELECTROMAGNETIC IMMUNITY
The equipment known as "LED Source" was designed to be used in the electromagnetic environment specified below.
Users must ensure that it is actually used in this environment.
Immunity trial
Electrostatic discharges
EN 61000-4-2
Fast transient bursts
EN 61000-4-4
Electrical shocks
EN 61000-4-5
Common mode ± 2 kV
Voltage dips, short
interruptions and
`voltage variations
EN 61000-4-11
Magnetic field with the
network frequency
(50/60 Hz)
NOTE: Ur is the power voltage nominal value applied during the trial.
011563 C - 11/2013
E N G L I S H
Compliance
Group 1
Class B
Complies
Complies
CEI 60601
Compliance
severity level
± 6 kV via contact
± 8 kV via the air
± 2 kV electric lines
± 6 kV via contact
± 1 kV finput/output lines
± 8 kV via air
Differential mode ± 1
± 6 kV via contact
kV
± 8 kV via air
<5% UT – for 10 ms
40% UT – for 100 ms
± 6 kV via contact
70% UT – for 500 ms
± 8 kV via air
<5% UT – for 5 s
± 6 kV via contact
3 A/m
± 8 kV via air
12
Electromagnetic environment - Guide
This product "LED Source" only uses radio-frequency energy for its
subsystems. Therefore its RF emissions are very low and are unlikely to
cause interference with nearby electronic devices.
This product "LED Source" may be used in all domestic settings,
including those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Electromagnetic environment - Guide
level
The floor should be wooden, concrete or tile. If the
± 6 kV
floor is covered with a synthetic material, the
± 8 kV
relative humidity should be at least 30%.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
If the user of this product requires it to continue operating
during main power supply interruptions, it is recommended
that this product be supplied by UPS or battery.
The magnetic field with the network frequency should be
at a characteristic level of a location (50 / 60 Hz) in a
traditional commercial or hospital environment.
15