GUIDE AND DECLARATION OF THE MANUFACTURER - ELECTROMAGNETIC EMISSIONS
The equipment known as "LED Source" is intended to be used in the electromagnetic environment specified below.
Users should make sure it is used in this environment.
Safety trial
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
NOTE 1 : At 80 MHz and 800 MHz, the higher frequency band should be used.
NOTE 2 : These recommendations may not apply in every situation. Electromagnetic wave propagation is modified by the
absorption and reflection due to structures, objects and persons.
For transmitters whose maximum power output is not listed above, the recommended separation distance d in metres (m) can be
determined by using the equation applicable to the transmitter frequency, where P is the maximum power output of the transmitter
in watts (W) as rated by the transmitter manufacturer.
16
E N G L I S H
CEI 60601
Compliance
severity level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Electromagnetic environment - Guide
level
Portable and mobile RF communication equipment should
not be used at a distance from this product, including
cables, that is less than the recommended separation
distance, calculated by applying the formula that
corresponds to the transmitter frequency.
d = 1,16
P
3V
P 80 MHz to 800 MHz
d = 1,16
P 800 MHz to 2,5 GHz
d = 2,33
3V/m
where
is the maximum output of the transmitter in watts
P
(W) rated by the transmitter manufacturer and (d) is the
recommended separation distance in metres (m).
The field levels emitted by the fixed RF transmitters –
which must be determined by an electromagnetic
measurement in situ – should be lower than the
compliance level in each frequency band.
Interference may occur in the vicinity of devices bearing
the following symbol:
011563 C - 11/2013