Operation; Switching The Device On And Off; Switching On The Device; Switching Off The Device - Seca 787 Instrucciones De Uso Y Garantía

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5. OPERATION

5.1 Switching the device on and off

Switching on the device

Switching off the device

5.2 Switching units of measurement (variant-dependent)

5.3 Measuring weight

28 •
1. Ensure that there is no load on the measuring station.
2. Press the on/off key.
All the elements of the display are shown briefly, then
display.
The measuring station is operational when
► Press the on/off key.
NOTE
In battery operation, the measuring station remains on standby for 30
seconds (factory setting). If there is no input or no load is put on the
weighing platform during standby time, the measuring station
switches off automatically.
CAUTION!
Patient hazard
In order to avoid misinterpretations, measuring results for medical use
must be displayed and used in SI units (weight: kilogrammes, length:
metres) only. Some devices offer the ability to display measuring
results in other units. This is only an additional function.
► Use the results exclusively in SI units.
► The use of measuring results in non-SI units is the sole responsibility
of the user.
You can switch between the metric units kilograms and centimeters and the
imperial units pounds and feet/inches during the measurement procedure.
The units of weight and height change at the same time.
1. Switch on the measuring station (➔ "Switching on the device", page 28).
2. Keep the
changes.

Weighing

CAUTION!
Injury from incorrect position of the measuring flap
Injuries may result if the measuring flap is at patient height when
folded up.
► Ensure that the measuring flap is above patient height before the
patient steps onto the weighing platform.
key depressed until the unit of measurement display
kg/lbs
appears in the
seca
is displayed.
0.0