4/ SPECIFICATIONS

4.1/ CLASS OF SAFETY

The C.A 61, 63 and 65 conform to standard IEC 1010 50 V CAT III - POL 2:
- installation category III
- degree of pollution 2 (environment without conducting pollution, except by temporary and
occasional condensation).
- rated voltage 50 V
Instruments entirely protected by double insulation.

4.2/ GENERAL SPECIFICATIONS

C.A 61
C.A 63 and 65
Power supply
battery 9 V (type 6 LR 61 or 6 LF 22)
Battery life
> 100 h (emission on infinite load)
> 50 h
(with battery 6 LR 61)
> 10 h (emission 20 mA at 10 V)
Watertightness
IP 52
IP 50
Dimensions
215 x 70 x 39 mm
Weight
310 g

4.3/ METROLOGICAL SPECIFICATIONS

The accuracies, given below, are expressed in ±(X% rdg + C), where rdg is the value read and
C a constant (expressed in physical units or in number of units of last digit). These accuracies
correspond to measurements made in the reference conditions:
- Temperature: 23°C ± 1 °C - Relative humidity: 45...75% RH

4.3.1 Receive mode

Measurement rate of each instrument 2 measurements per second
C.A 61
Measurement of DC voltage
Range
Resolution
Accuracy over 1 year
Input resistance
30 V
0.01 V
±(0.2% rdg + 2)
1 MW
Measurement extent : -5.00...30.00 V
Max permitted voltage: 60 V DC or AC peak
Series mode rejection: ³ 60 dB at 50 and 60 Hz
Max permitted voltage in common mode: 60 V AC or 85 V peak
Common mode rejection: ³ 120 dB at 50 and 60 Hz
Temperature coefficient: £ 10% of the accuracy/°C
45