EN-4
Input
The scale will respond to various commands sent via the interface.
Command
Legacy Command (2)
IP
P
CP
CA
SP
xS
xP
xA
Z
T
xT
PU
xU
xM
PV
V
\EscR
Notes:
Auto-Print Operation
Once Auto-Print is activated in the menu, the balance will send data as required. To temporarily stop
Auto-Printing, press the PRINT key. If there is data in the print buffer the printer will finish printing this
data. A second press will resume Auto-Printing.
(Note: In some modes, stopping or resuming Auto-Print must be done when the balance is not
reading "0".)
ACCESSORIES
For a complete listing of OHAUS printers and other accessories, contact OHAUS Corporation or visit
www.OHAUS.com
Function
Immediate Print of displayed weight (stable or unstable).
Print displayed weight (stable or unstable).
Continuous Print.
Print on Stability.
0S: Turn off "Stable Only" menu item and allow unstable
print. 1S: Turn on "Stable Only" menu item and only print
stable print.
Interval Print x = Print Interval (1-3600 sec), 0P turns auto
print OFF
Same as pressing Zero Key
Same as pressing Tare Key.
Download Tare value in grams (positive values only).
Sending 0T clears tare (if allowed).
Print current unit: g, kg, lb, PCS
Set scale to unit x: 1=g, 2=kg, 3=lb,
Set scale to mode x. M will scroll to next enabled mode.
Version: print name, software revision
Global reset to reset all menu settings to the original
factory defaults
The second commands listed are "legacy' commands,
which maintain compatibility with older products.
RS232 Interface