Instructions for CEREC Bluecam and CEREC Omnicam
62 82 839 D3492
D3492.201.02.06.02    10.2018
2.7 High-level disinfection of the mirror sleeve with the HLD set 66 05 120
9. Insert the metal bracket with the metal sleeve.
10. Place the cover on the container and leave the mirror sleeve in the
disinfectant for at least 12 minutes (CIDEX
will not be damaged if it remains in the disinfectant for longer than
the required time, but removing it shortly after the required time is
recommended.
2 CEREC Omnicam
®
OPA). The mirror sleeve
9