2.7 High-level disinfection of the mirror sleeve with the HLD set 66 05 120

High-level disinfection of the mirror sleeve with
2.7
the HLD set 66 05 120
NOTE
The hot air sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection (HLD) is as follows –
assuming the CIDEX
your dealer:
A
HLD tank
B
Cover
1. Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve  [→ 6]").
2. Always use personal protective equipment when implementing the
high-level disinfection if not before.
3. Remove the mirror sleeve from the CEREC Omnicam (see
"Removing the mirror sleeve  [→ 5]").
4. Place the white protective cap on the camera head and place the
camera in the camera storage unit.
5. Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.
®
OPA disinfectant is available in your country via
C
Metal bracket
D
Plugs
2 CEREC Omnicam
7