2.7 High-level disinfection of the mirror sleeve with the HLD set 66 05 120
10
Instructions for CEREC Bluecam and CEREC Omnicam
11. Remove the cover and take the bracket with the mirror sleeve out of
the container.
12. Carefully rinse the mirror sleeve with tap water for at least 30
seconds while holding it with the bracket.
13. Remove the bracket from the mirror sleeve carefully and slowly by
pushing the bracket downwards. Ensure that the mirror sleeve is
pointing upwards so that no fluids are able to penetrate into the
interior of the mirror sleeve.
14. Remove the plug from the bracket.
15. Dry the plug.
16. Dispose of the fluid and store the bracket in the empty container if
the HLD set is not used for more than one week.