Use Of Cerec Omnicam Disposable Sleeves - Dentsply Sirona CEREC Bluecam Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

2 CEREC Omnicam

2.10 Use of CEREC Omnicam disposable sleeves

14
Use of CEREC Omnicam disposable sleeves
2.10
The CEREC Omnicam can be operated with disposable plastic sleeves
in order to ensure maximum infection control. The sleeves are available
using order number 66 32 264.
NOTE
The disposable sleeves cannot be used in the following
circumstances:
- with the CEREC AF and CEREC AI system,
- when using color analysis,
- when using CEREC Ortho SW software.
1. Remove the disposable sleeve from the CEREC Omnicam
reflective sleeve after each patient and dispose of this following
standard procedures.
2. Disinfect the camera with a wipe (see "Wipe disinfection for the
camera and mirror sleeve  [→ 6]").
3. Remove one disposable sleeve from its packaging. Slide the sleeve
onto the CEREC Omnicam reflective sleeve until it stops so that it is
firmly held on the camera. Hold the camera in other places, not just
on the sleeve in order to prevent it from falling.
4. Place the camera in the tray so that it can warm up for at least 15
minutes before intraoral use. Ensure that the tray is disinfected (see
"Non-critical surfaces with no monitor").
5. When using the intraoral camera with disposable sleeves, the
camera should be positioned closer to the teeth than it would be
without a sleeve. If the 3D preview indicates artifacts, check
whether the window of the disposable sleeve is clean and wipe it
down with alcohol.