Aesculap
Lektrafuse HF generator GN200
5.5
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
Unplug the device before cleaning.
DANGER
Do not use flammable or explosive cleaning or
disinfecting solutions.
Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
Only clean and disinfect the product manually.
CAUTION
Do not sterilize the product under any circum-
stances.
Damage to the product due to inappropriate clean-
ing/disinfecting agents!
Only use cleaning/disinfecting agents approved
CAUTION
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
Use cleaning agents and disinfectants according
to the manufacturer's instructions.
Observe specifications regarding concentration,
temperature and exposure time.
Validated cleaning and disinfection procedure
Validated procedure
Special features Reference
Wipe disinfection for
–
electrical devices with-
out sterilization
12
®
Chapter Wipe disinfec-
tion for electrical devices
without sterilization