Failure Messages; Problem Solving - B.Braun Aesculap GN200 Instrucciones De Manejo/Descripción Técnica

Ocultar thumbs Ver también para Aesculap GN200:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
7.4

Failure messages

A failure is indicated as follows:
Clear sound sequence (as necessary, may interrupt the sound of the HF
discharge)
The red error light 6 on the front side of the HF generator lights up
Error message in the display 5 on the front side of the HF generator
Failure code
Cause
F001
Software error on the HF generator
F002
Dosage error of HF generator
F003
Hardware error
F004
Internal communication error of the HF generator
F005
Overheating of HF generator
F006
Error in the operating mode selection button (MODE) 2 (activated
when switching on)
F007
Error with HF energy output
F008
Fault: HF outside the calibration
F009
Fault: Internal power supply
7.5

Problem solving

Finding
HF generator does not switch on
The device is switched on, but no displays are illumi-
nated and the self-test is not being carried out
Self-test fails
HF generator is active (power ON signal lamp 9 lights
up), and accessories are connected, but there is no HF
discharge
Note
If the failure occurs during HF discharge, the vessel sealing may be incom-
plete or insufficient.
In order to ensure that sufficient hemostasis has been achieved, follow the
procedure for regrasp errors, see Regrasp error.
Note
If the failure continues after turning the device off and on again twice, con-
tact your national B. Braun/Aesculap agency, see Technical Service.
Remedy
Ensure that the power cord has been correctly plugged into the power supply 13 of the HF
generator.
Ensure that the power cord is connected to a functioning power source.
Ensure that the power OFF/ON switch 12 of the HF generator is in the power ON position.
Check the fuse in the fuse box and replace if necessary, see Fuse replacement.
If the error continues, replace power cord.
Switch off HF generator, wait 5 s, switch HF generator on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Ser-
vice.
Switch off HF generator, wait 5 s, switch HF generator on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Ser-
vice.
Verify that the required foot switch (if used ) is properly connected and plugged in.
Ensure that the Caiman Seal & Cut Instrument is correctly connected and plugged in (green
ring 1 is lit up).
In order to start the HF discharge, activate the button on the instrument or the foot switch.
If the error continues, replace Caiman instrument.
Remedy
Switch off HF generator, wait 5 s, switch HF generator on
Switch off HF generator, wait 5 s, switch HF generator on
Switch off HF generator, wait 5 s, switch HF generator on
Switch off HF generator, wait 5 s, switch HF generator on
Switch off HF generator and leave to cool down
Check the air stream around the generator
Switch off HF generator, wait 5 s, release operating mode
selection button (MODE) 2, switch HF generator on
Switch off HF generator, wait 5 s, switch HF generator on
Switch off HF generator, wait 5 s, switch HF generator on
Switch off HF generator, wait 5 s, switch HF generator on
17

Publicidad

Tabla de contenido
loading

Tabla de contenido