Error Code - Coltene CanalPro CL Instrucciones Para El Uso

Ocultar thumbs Ver también para CanalPro CL:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
„ Changing Batteries
Prepare small screwdrivers
1. Turn off the power of the motor handpiece.
2. Remove the rubber cover from the battery cover with
a thin flatblade screwdriver. (Fig. 13)
3. Remove the screw fixing the battery cover with a
screwdriver. (Fig. 14)
4. Remove the battery cover by sliding it toward the
charging terminal. (Fig. 15)
5. Remove old batteries.
6. Insert new batteries according to the polarity mark-
ing in the battery box. A mix-up between the plus
and minus sides will not allow operation of the hand-
piece.
7. Attach the battery cover.
8. Tighten the screw with a screwdriver. Do not over-
tighten it.
9. Insert the rubber cover back into the screw hole of
the battery cover in the originally inserted direction.
Changing of batteries is now completed. Charge them
fully before use.
CAUTION
- Do not misplace the rubber cover and screw.
- The used nickel metal hydride batteries are recyclable, but their disposal may sometimes not be permit-
ted by the respective country. Return them to your dealer.
46
Rubber Cover
Battery Cover
Fig. 13
Screw
Battery Cover
Fig. 14
Battery Cover
Fig. 15

12. Error Code

If the motor handpiece stops due to an abnormality such as a malfunction, overload, break or wrong use, it
automatically checks the state of the control unit and detects the cause of the abnormality and displays an
error code on the LCD panel. If an error code is displayed, turn on the power again and check whether the
same error code is displayed. If the same error code is displayed, take action by referring to the instructions
provided in the "Check/Remedy" column in the following table.
Error code Error
During rota-
Self-Check error
tion of the
Overcurrent
motor hand-
piece
Overvoltage
Overheating of
motor
At the time
Charger failure
of Charging
Low voltage of
batteries
High voltage of
batteries
Outside the
range of work-
ing temperature
Other
Abnormal heat
generation from
batteries
At the time of
Beyond the up-
calibration
per limit
Below the lower
limit
Cause
Check/Remedy
Malfunction of circuit
Contact your dealer
The motor handpiece is
Remove load
locked. (at the time of
the auto reverse mode)
Malfunction of circuit
Contact your dealer
High load was continuous-
Leave it as it is until it
ly applied to the motor
cools down.
handpiece for a relatively
long time.
Malfunction of charger
Contact your dealer
The voltage of batteries is
Put the batteries into the
too low (The life of batter-
battery chamber, or replace
ies or not inserted.)
with new batteries.
The voltage of batteries is
Contact your dealer.
too high. (Malfunction of
circuit)
Outstanding the range of
Use within the range of
working temperature or
working temperature.
break in the thermistor in
the battery section
The batteries generate
Replace the batteries. If the
abnormal heat.
heat generates from the
new set of batteries, mal-
function of the circuit may
be suspected. Contact your
dealer.
The operating life of the
Replace the motor
motor handpiece or
handpiece or contra angle
contra angle has expired.
head.
47

Publicidad

Tabla de contenido
loading

Tabla de contenido