Guidance and manufacturer's declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Emissions Test
RF emission CISPR 11
RF emission CISPR 11
Harmonic emission
IEC/EN 61000-3-2
Voltage fluctuations
/ flicker emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Compliance
Group 1
Class B
Not applicable
Complies
IEC 60601-1
test level
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
+/- 2 kV for power
supply lines
+/- 1 kV for input /
output lines
+/- 1 kV line(s)
to line(s)
+/- 2 kV line(s)
to earth
<5% Ut (95% dip
in Ut) for 0.5 cycle
40% Ut (60% dip
in Ut) for 5 cycles
70% Ut (30% dip
in Ut) for 25 cycles
<5% Ut (95% dip
in Ut) for 5 sec.
Electromagnetic Environment Guidance
The system use RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The system is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Compliance
Electromagnetic Envi-
Level
ronment Guidance
+/- 8 kV contact
If floors are covered with
+/- 2, 4, 8, 15 kV
synthetic material the
air
relative humidity shall
be least 30 %.
+/- 2 kV power
Mains power quality
supply lines N/A
shall be that of a typical
commercial or hospital
environment.
Mains power quality
shall be that of a typical
commercial or hospital
environment.
100% reduction
Mains power quality
0.5 period
shall be that of a
typical commercial or
40% reduction
hospital environment.
for 5 periods
If the use of the system
requires continued
30% reduction
operation during power
for 25 periods
mains interruptions
the system can be
100% reduction
powered by the built in
for 5 sec.
rechargeable battery.
15