5-1 Preparation
1) Wear eye protection, a mask and gloves to avoid infection.
2) Disconnect the light probe from the motor.
5-2 Cleaning
■ Manual Cleaning (External Surface)
1) Remove debris on the light probe surface using a brush under running water (at 38 ºC or less). Do NOT use a wire brush.
2) Wipe off moisture on the surface using an absorbent cloth.
CAUTION
• If water enters the light probe, using an air syringe, blow air into the rear of the light probe to remove the internal water.
■ Automatic Cleaning (External Surface)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.