5-4 Disinfection
■ Manual Disinfection (External Surface)
Wipe the surface with cloth soaked in disinfectant solution such as alcohol.
■ Automatic Disinfection (External Surface)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.
5-5 Sterilization
Sterilize the product by autoclave sterilization. After each patient sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Type
Gravity-displacement Autoclave
Pre-vacuum Autoclave
3) The light probe should remain in the autoclave pouch until required for use.
8
Temperature
132°C
134°C
Time
more than 15 min.
more than 3 min.