Sterilization instructions

Prior to sterilization:
Clean and disinfect contra-angle handpiece (6.5).
Protect contra-angle handpiece by applying T1 Spray (6.3).
P
Contra-angle handpieces may be sterilized only in the steam
Professional
DAC
sterilizer with saturated water vapor at 134 °C (274 °F),
3 min. holding time (2.04 bar; 29.59 psi overpressure) .
Steam sterilizers that fulfill the requirements of EN 13060,
class B or S and are suitable for the sterilization of contra-
angle handpieces:
e.g. DAC PROFESSIONAL by Sirona.
The Hygiene Division of Sirona
The Hygiene Division of Sirona
ATTENTION! Do not exceed 140 °C (284 °F) even dur-
ing the drying phase!
6 Step-by-step Instructions
21