CLEANING AND DISINFECTING YOUR J3
DRY FLOATATION® INSERT
Cleaning and disinfecting are two separate processes. Cleaning
must precede disinfection.
CLEANING THE DRY FLOATATION® INSERT
To clean the DRY FLOATATION
recess in the foam base, taking care to gently pull the valve through
the foam. Deflate the cushion, close the valve(s) and place in a
large sink. Gently scrub the air cells and the spaces between them
with a soft plastic bristle brush, sponge or washcloth. Liquid dish-
washing or laundry detergent or other general-purpose cleaning
product is suggested. For institutional cleaning, a general house-
keeping disinfectant cleaner may be used.
DISINFECTING THE DRY FLOATATION® INSERT
To disinfect the DRY FLOATATION
uct that meets infection control needs. A disinfectant with tubercu-
locidal properties is recommended. Follow the manufacturer's dilu-
tion directions. For home disinfection, use solutions of approxi-
mately 1/2 cup liquid household bleach per quart of water (125 ml
bleach to 1 liter of water). Keep the clean cushion wet with bleach
solution for 10 minutes. Rinse thoroughly and dry.
Disinfectants are not effective on porous surfaces, such
as foam, if the cushion becomes soiled or contaminated.
DO NOT use with multiple users.
CAUTION– DO NOT USE OIL-BASED LOTIONS OR LANOLIN ON YOUR
CUSHION AS THEY MAY COMPROMISE THE INTEGRITY OF
THE NEOPRENE RUBBER.
®
insert, gently pull it out of the
®
Insert, use a germicidal prod-
WARNING
C L E A N I N G A N D M A I N T E NA N C E
REPAIRS
Each J3 Cushion with a Dry Floation® insert is shipped complete with
a repair kit and instructions for minor repairs. The following informa-
tion can be used if repair instructions have been lost or misplaced.
If A Leak Occurs
1. Place your cushion cells down and over-inflate until all four
corners touch.
2. Immerse the over-inflated cushion in water to locate the leak.
3. Once the leak has been located, mark the leak by inserting a
toothpick into the hole
4. Allow the cushion to dry thoroughly.
5. Remove the toothpick, and clean the area around the hole
with the alcohol wipe provided. Let the surface dry.
6. Peel the backing from the patch provided and, centering the
patch, place it over the hole.
7. Press the patch down firmly until a good seal is achieved.
8. The cushion is now ready for use. Adjust the cushion for
proper inflation following the instructions in this manual.
Although patches do not change the performance of the cushion,
they should be monitored periodically to ensure effectiveness in
addressing the leak. If the patch is not successful in addressing the
problem, or if more extensive damage is present, please contact
your supplier immediately.
Replacing Valves
1. Firmly grip the rubber valve stem and remove the old inflation
valve by using pliers.
CAUTION– DO NOT REMOVE AN INFLATION VALVE BY HOLDING THE
CUSHION AND PULLING THE VALVE, DOING SO MAY TEAR
THE CUSHION.
2. Moisten the ribbed end of the replacement valve and insert it
into the valve stem until the rubber is snug to the body of
the valve.
5
115344 Rev. A