7.3

Microscope assembling

7.3.1 B-1000POL
1. Insert the Bertrand lens ① on the frame and
lock the locking screw ② with the provided Allen
wrench. (Fig. 1)
2. Insert the optical head on the Bertrand lens and
lock the locking screw with the provided Allen
wrench. (Fig. 2)
Hold the head with one hand during the
locking in order to avoid that the head falls.
3. Insert both eyepieces into the tubes of the optical
head. (Fig. 3)
One of the two eyepieces is equipped with
a crosshair for centering the entire optical
system. It is advisable to insert the eyepiece
with crosshair in the right eyepiece sleeve.
4. Insert the condenser under the stage: position
until it is well inserted into its holder (under the
condenser there is a pin that must fully enter the
holder guide). (Fig. 4)
5. Lock the condenser fixing knob ①.