13. Microphotography

13.1

Use of C-mount cameras

1. Loosen the clamping screw ① on the trinocular
port and remove the dust cap ②. (Fig. 56)
2. Screw the C-mount adapter ③ to the camera ④
and insert the round dovetail of the C-mount into
the empty hole of the trinocular port, then tighten
the clamping screw ①. (Fig. 57)
13.2 Use of Reflex cameras
1. Insert the Reflex adapter ① into the relay tube to
the microscope ②.
2. Screw the "T2" ring ③ (not provided) to the reflex
adapter.
3. Connect the Reflex camera ④ to the "T2" just
installed (Fig. 58).
4. Mount the other end of the relay tube ② into the
empty hole of the trinocular port, then tighten the
clamping screw. (Fig. 56)
"T2" ring is not provided with the microscope, but
is commercially available.
While shooting dark specimens, darken eyepieces
and viewfinder with a dark cloth to minimize the
diffused light.
To calculate the magnification of the camera:
objective magnification * camera magnification *
lens magnification.
If using an SLR camera, mirror movement may
cause the camera to vibrate.
We suggest lifting the mirror, using long
exposure times and a remote cord.