Dräger Idmed ToFscan Manual Del Usuario página 25

Monitor de relajación muscular
Tabla de contenido

Publicidad

Surges
Line-to-line
61000-4-5
Surges
Line-to-ground
61000-4-5
Conducted
disturbances
induced by RF
61000-4-6
fields
RATED
power
frequency
61000-4-8
magnetic fields
Voltage dips
61000-4-11
Voltage
interruptions
61000-4-11
Estimulaciones
• TOF (Train Of Four), cálculo del T4/T1 y del T4/Tref.
• TOF Auto (ToF programado de 15 s a 15 min).
• TET (Tétanos 50 Hz)
• DBS (Double Burst Estimulación) modo 3.3 y 3.2
• PTC (Post Tetanic Count)
• ATP (Modo automático TOF y PTC)
• TWITCH (Single Twitch) 0,1 Hz y 1 Hz.
Sensor de aceleración
• Acelerómetro tridimensional (+/- 8 G - 10 bits, Frecuencia : 200 Hz, Resolución 0.016G)
Estimulación eléctrica
• Corriente de salida constante de 0 a 60 mA (precisión +/- 10 %)(bajo carga resistiva de
4Kohms)
• Monofásica, duración del impulso 200 µs, frecuencia 50 Hz
• Electrodos de estimulación o electrodos de ECG:
- Capaz de soportar hasta 300 voltios con una corriente de 60 mA.
- La superficie de contacto debe ser superior a 1,8 cm².
Ejemplos de electrodos recomendados:
- Electrodos RED DOT ref.2560 de la empresa 3M
- Electrodos F9047 de la empresa FIAB
Manual del Usuario ToFscan
IEC
± 0,5 kV, ± 1 kV
± 0,5 kV, ±
IEC
1 kV,
± 2 kV
3 V
0,15 MHz –
80 MHz
6 V in ISM
IEC
bands
between 0,15
MHz and
80 MHz
80 % AM at
1 kHz
IEC
30 A/m
50 Hz or 60 Hz
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
IEC
0 % UT;
1 cycle and
70 % UT; 25/30
cycles
Single phase:
at 0°
IEC
Mains power quality should be that of
± 0,5 kV, ± 1 kV
a typical residential, commercial or
hospital environment.
Mains power quality should be that of
± 0,5 kV, ± 1
a typical residential, commercial or
kV, ± 2 kV
hospital environment.
3 V
0,15 MHz –
80 MHz
6 V in ISM
In order to prevent electromagnetic
bands
disturbance, keep minimum separa-
between 0,15
tion from RF communication equip-
MHz and
ment of 30cm
80 MHz
80 % AM at
1 kHz
Mains power quality should be that of
30 A/m
a typical residential, commercial or
50 Hz or 60 Hz
hospital environment
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Mains power quality should be that of
a typical residential, commercial or
0 % UT;
hospital environment.
1 cycle and
70 % UT; 25/30
cycles
Single phase:
at 0°
Mains power quality should be that of
a typical residential, commercial or
hospital environment.
25

Publicidad

Tabla de contenido
loading

Tabla de contenido