Troubleshooting
Problem
Unit does not power ON when
ON/OFF button at console is
pressed.
Unit does not power Off when
ON/OFF button at console is
pressed.
Irrigation Pump does not
rotate.
Noise is heard from console even
though there are no procedures
being performed.
This noise was not heard when
the console was first powered up.
Excessive handpiece noise.
Lack of power to console.
Handpiece vibration or
extremely hot.
Excessive vibration of
dissection tool.
Unable to place or lock
attachment.
Unable to remove attachment.
Attachment is hot.
Craniotome attachment is bent. 1. Excessive force used in operation.
Inoperative Foot Control.
10
Troubleshooting Chart
Possible Cause
1. Not enough time for console to
detect signal.
1. Not enough time for console to
detect signal.
1. Irrigation pump is not enabled on
display.
1. The air pump continues to
functioning even when the handpiece
is not being run.
1. Faulty internal component.
1. Plug is not fully inserted/power
switch not turned on.
2. Outlet is not functional.
1. Faulty internal component.
2. Internal motor overheats due to
continuous extreme cutting at near
stalling conditions.
3. Hose may be kinked.
4. Handpiece out of balance.
5. Handpiece hose damaged.
1. Dissection tool may be bent.
2. Dissection tool may not be fully
seated.
3. Improper attachment and dissection
tool combination.
4. Possible attachment bearing
damage.
1. The Indicator Bar in the handpiece
in not aligned with the Unlock (Load)
symbol.
2. Misalignment of the Angle
Attachment shaft.
1. Attachment has safety locking
mechanism.
2. Attachment is captured by dissection
tool.
1. Debris lodged inside attachment.
2. Possible bearing damage.
1. Connector may not be fully inserted. 1. Verify plugs for handpiece and foot
Possible Solution
1. Keep the ON/OFF button pressed for
at least 1 second.
1. Keep the ON/OFF button pressed for
at least 2 seconds.
1. Keep the Irrigation button Pressed for
at least 1 second.
1. This is a normal condition, the air
pump will turn off automatically after
approximately 15 minutes.
1. Return attachment to DePuy Synthes
Power Tools for service.
1. Verify plugs to wall and back of
console are pushed in completely.
2. Verify another piece of electrical
equipment can receive power from outlet.
1. Return to DePuy Synthes Power Tools
for service.
2. Decrease cutting force, allow motor
to cool.
3. Unkink hose (verify it is not being
pinched or clamped to table).
4. Return to DePuy Synthes Power Tools
for service.
5. Return to DePuy Synthes Power Tools
for service.
1. Replace with new dissection tool.
2. Reassemble dissection tool and
attachment.
3. Only use correct dissection tool with
appropriate attachment.
4. Return to DePuy Synthes Power Tools
for service.
1. Manually rotate the Indicator Bar to the
Unlock (Load) Symbol.
2. Manually rotate the unit shaft.
1. Follow Instruction carefully to remove
the attachment.
2. Manually rotate the Indicator Bar to the
Unlock (Load) Symbol.
1. Clean attachment using Anspach
Attachment Cleaning Instructions..
2. Return attachment to DePuy Synthes
Power Tools for service.
1. Do not use, replace attachment.
control are fully inserted. Plug will latch in
place once fully inserted.