Trouble Shooting

Problem
Possible Cause
Excessive handpiece
1. Faulty internal component.
noise.
Lack of power to
1. Plug is not fully inserted/power switch not
console.
2. Outlet is not functional.
Handpiece vibration
1. Faulty internal component.
or extremely hot.
2. Internal motor overheats due to continuous
3. Hose may be kinked.
4. Handpiece out of balance.
5. Handpiece hose damaged.
Excessive vibration
1. Dissection tool may be bent.
of dissection tool.
2. Dissection tool may not be fully seated.
3. Improper attachment and dissection tool
4. Possible attachment bearing damage.
Attachment is hot.
1. Debris lodged inside attachment.
2. Possible bearing damage.
Craniotome
1. Excessive force used in operation.
attachment is bent.
Inoperative Foot
1. Selected foot control on face panel does not
Control (LEDs are
illuminated on face
2. System is set in hand control mode (LED
panel)
3. Plug may not be fully inserted
4. Handpiece in "Safe" position or cutter is
5. Defective attachment
Inoperative Hand
1. Defective internal component
Control
2. Hand Control may not be fully installed.
3. Selected Hand Control on face panel not
4. Internal handpiece switch activating Hand
Trouble Shooting Chart
turned on.
extreme cutting at near stalling conditions.
combination.
match foot control in use
light is illuminated)
not rotating
in use.
Control may be damaged.
Possible Solution
1. Return attachment to DePuy Synthes
Power Tools for service.
1. Verify plugs to wall and back of console are
pushed in completely.
2. Verify another piece of electrical equipment
can receive power from outlet.
1. Return to DePuy Synthes Power Tools for
service.
2. Decrease cutting force, allow motor to cool.
3. Unkink hose (verify it is not being pinched
or clamped to table).
4. Return to DePuy Synthes Power Tools for
service.
5. Return to DePuy Synthes Power Tools for
service.
1. Replace with new dissection tool.
2. Reassemble dissection tool and attachment.
3. Only use correct dissection tool with
appropriate attachment.
4. Return to DePuy Synthes Power Tools for
service.
1. Clean attachment using Anspach
Attachment Cleaning Instructions.
2. Return attachment to DePuy Synthes
Power Tools for service.
1. Do not use, replace attachment.
1. Press Foot Control button on face panel to
select other Foot Control
2. Press Foot Control button on face panel
to set it for Foot Control 1 or 2, depending
which one is in use.
3. Verify plug for handpiece and foot control
are fully inserted. Plug will latch in place
once fully inserted.
4. Verify cutting burr is completely engaged
and handpiece's knurled knob is in secure
position.
5. Clean attachment or replace attachment.
1. Return hand control to DePuy Synthes
Power Tools for service.
2. The circumferential groove on handpiece
housing will be exposed when the Hand
Control is fully installed.
3. Press Hand Control button on face panel to
illuminate Hand Control LED
4. Return handpiece to DePuy Synthes Power
Tools for service.
9