Mettler Toledo SevenGo pro SG6 Instrucciones De Manejo página 17

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 77
Set temp.
Set MTC temperature
If the meter does not detect a temperature probe, it automatically switches to the manual temperature com-
pensation mode and MTC appears. In this case the temperature of the sample has to be entered in the Set MTC
menu (0 °C...60 °C). The meter calculates the temperature-adjusted electrode slope using this temperature
and shows the temperature-compensated value in the measurement display.
Note
For better accuracy, we recommend the use of either a built-in or a separate temperature probe. If a tempera-
ture probe is used, the ATC symbol and the sample temperature are displayed.
Set meas.
EP format
With this menu you can choose between one of three different types of endpoint formats:
"Auto"
Automatic endpoint
is a special algorithm that determines the end of an individual reading, depending
on the behavior of the sensor used.
Stability criterion for dissolved oxygen measurements
The meter will endpoint if the signal changes less than 0.052 nA in 30 seconds.
"Manual"
Manual endpoint
means the meter never endpoints the reading unless the user manually presses the
Read key.
"Timed"
With the timed endpoint
Note
Every measurement can be endpointed manually by pressing the Read key. The meter then displays
The table below shows how the entpoint format is displayed in the course of the measurement.
Preselected format
Auto endpoint
Manual endpoint
Timed endpoint
) The actual enpoint format (last column) not the preselected is stored with the data.
1
METTLER TOLEDO SevenGo pro™ dissolved oxygen
, the reading is ended automatically when the set time period has elapsed.
Start of measurement
Read
Read
Read
Signal stability
Read
Menus
.
Endpointed measure-
ment
)
1
13

Publicidad

Tabla de contenido
loading

Tabla de contenido