WIKA DIH10 Manual De Instrucciones página 18

Ocultar thumbs Ver también para DIH10:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Commissioning, operation
6.3 Programming the display
6.3.1 Programming table for programming the display
GB
PN
Description
0
Calibration mode
0 = Sensor calibration (signal must be
applied, only for factory setting)
1 = Rescaling (display value at 4 ... 20 mA)
1
Full scale value (programming the display
value at 20 mA, e.g. 600)
2
Initial value (programming the display value
at 4 mA, e.g. 100)
3
Decimal point or unit selection
(for programming a unit, the value shifts to
the left)
4
Display and averaging time in 1/10 of a second 5.0 ... 100.0 10
5 **
Zero point stabilisation (the ± range in which
0000 will be displayed)
50 *** Definition of PIN code > 0000 (for
programming lock)
51 *** Authentication by PIN code (removal of
programming lock)
100
Number of reference points (reference points
for sensor calibration, reference points
reduce the measuring rate)
101 ...
Reference points (as programmed under
130
PN100)
200
TAG number
*
For factory setting
** For programming a value >1 a hysteresis of 0.1 % will be activated. This prevents
the display jumping.
*** optional
(If no PIN code has been defined, PN50/PN51 are not available. A PIN code can
only be initially programmed via the serial interface for the factory programming. If a
PIN code has been defined (display after segment test: Pin), the programming lock
must be released, using the PIN code defined via PN50, through PN51 under input,
as otherwise, any changes made under the other PN's will not be able to be saved.
In order to reactivate the PIN code via PN50 on completion of any programming,
the programming mode must be left (by the input of 0000 under PN50 or by briefly
interrupting the current loop).
18
Range
0/1
-999...9999 150
-999...9999 0
0 / 0.0 / 0.00
/ 0.000 /
°F / °C
0 ... 100
0000...9999 0000
0000...9999 0000
0 ... 30
-999 ... 9999 0
0000 ... 9999 0
WIKA operating instructions model DIH10
Factory
default *
1
°C
2
0

Publicidad

Tabla de contenido
loading

Tabla de contenido