4 SPARES
Probe caps wear over time and the gauge will inform the user during
the calibration procedure if the cap needs replacing . When tested
using smooth surfaces, probe caps have been scanned in excess of
50km (30 miles). Replacement caps are available to purchase from
Elcometer or your local supplier.
Description
Scan Probe Caps (3 per pack)
Foil Set; Scale 1; 0 - 1500µm (0 - 60mils)
Foil Set; Scale 1; 0 - 1500µm (0 - 60mils) - Certified
Foil Set; Scale 2; 0 - 5mm (0 - 200mils)
Foil Set; Scale 2; 0 - 5mm (0 - 200mils) - Certified
Ferrous Zero Test Plate, 1%
Non-Ferrous Zero Test Plate, 1%
5 IMPORTANT INFORMATION
"Scan" and "Auto Repeat" modes are only available on Elcometer 456
Model T separate gauges. The probe cap must be fitted to the Scan
Probe when used in these modes to avoid damage to the probe tip.
Use in these modes without the cap will invalidate the probe warranty.
Scan Probes can be used in "Standard" mode - with or without the
probe cap fitted - when connected to any Elcometer 456 separate
gauge.
When fitted with the probe cap, Scan Probes are not recommended for
use on coating thicknesses below 100µm (4mils).
h
When the cap thickness falls below 200µm (8mils).
en-5
R
h
Part Number
T456C23956
T99022255-1
T99022255-1C
T99022255-2
T99022255-2C
T9994910-
T9994911-
www.elcometer.com