5. Cleaning and daily disinfecting
6.
After cleaning and disinfecting the parts, rinse them thoroughly with clean
water and allow the parts to air dry in a clean environment.
7.
If the compressor is dirty, wipe it clean with a soft
cloth moistened with water or mild detergent.
Caution: The compressor is not
8.
If the outside of the air tube is dirty, wipe it clean
with a soft cloth moistened with water or mild detergent.
9.
Reassemble the nebuliser kit. Put the parts in a dry, sealed bag and place
in the storage pocket of the carrying bag.
Note: Nebuliser kit should be replaced after using for 6 months.
5.2

Disinfecting the nebulising parts

Always clean before disinfection. (See Section 5.1)
Always disinfect the medication tank, mouthpiece and other nebulising parts after
the last treatment of the day. To disinfect the parts, use one of the methods
described below:
A. Use a commercially available disinfectant. Follow the instructions provided by
the disinfectant manufacturer.
1.
Submerge the parts in the cleansing solution for the specified period.
2.
Remove the parts and discard the solution.
3.
Rinse the parts with clean hot tap water, shake off excess water and allow
to air dry in a clean environment.
B. Parts may be boiled between 15 to 20 minutes (except for PVC mask).
After boiling, carefully remove the parts, shake off excess water and allow to air dry
in a clean environment.
20
waterproof.