Directions For Use - Masimo RD SedLin Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
RD SedLine® Patient Cable
Reusable
Prior to using this Cable, the user should read and understand the Operator's Manual for the Device or Module, this
Directions for Use, and the Directions for Use for the Sensor.
DESCRIPTION
The RD SedLine® Patient Cable transfers analog EEG signals collected from the Sensor to the Module for processing. The
Patient Cable is a reusable cable and may be used from patient to patient.
Always refer to the operator's manual for complete instructions or additional instructions.
The RD SedLine Patient Cable is compatible with the RD SedLine EEG Sensor.
WARNINGS, CAUTIONS AND NOTES
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Masimo sensors and cables are designed for use with instruments licensed to use Masimo sensors.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Failure to properly connect the sensor or the module to the cable will result in intermittent readings, inaccurate
results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or
disconnecting either end.
• Always refer to the module operator's manual for complete instructions or additional instructions.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
• Caution: Replace the cable when a replace cable message is displayed, or when a low SIQ message is consistently
displayed while monitoring consecutive patients after completing the low SIQ troubleshooting steps identified in the
monitoring device operator's manual.
• Note: The cable is provided with X-Cal™ technology to minimize the risk of inaccurate readings and unanticipated loss
of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time. Replace the cable when
the patient monitoring time is exhausted. Some legacy monitoring devices do not distinguish between the cable and
the sensor. In these devices, a replace sensor or similar message could pertain either to the sensor or the cable.
INSTRUCTIONS
A) Connecting
To connect the Patient Cable to the Module
1. Align the ridged Patient Cable connector end with the available Module connector end.
2. Push to insert.
To connect the Patient Cable to the Sensor
1. Align the yellow indicator connector on the end of the RD SedLine Patient Cable with the yellow indicator on the Sensor
Connector.
2. Refer to Fig. 1. Push together until the connections fit securely.
B) Disconnecting the Sensor
To disconnect the Patient Cable from the Module
Gently pull the Patient Cable out of the Module by grasping the patient cable connector end.
To disconnect the Patient Cable from the Sensor
Gently pull the Sensor out of the Patient Cable connector from the Sensor's connector end.

DIRECTIONS FOR USE

LATEX
Not made with natural rubber latex
PCX-2108A
02/13
3
en
Non-sterile
9501E-eIFU-0518

Publicidad

Tabla de contenido
loading

Tabla de contenido