CLEANING
1. Remove the sensor from the patient and disconnect it from the patient cable.
2. Clean the surface of the cable by wiping it with a 70% isopropyl alcohol pad, quaternary ammonium chloride solution, or
a 10% (1:10) chlorine bleach to water solution.
3. Wipe all surfaces of the cable.
4. Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the cable.
5. Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.
The surface of the cable has been tested to be chemically resistant to the following solutions: Quaternary Ammonium
Chloride solution wipe, Accelerated Hydrogen Peroxide, Chlorhexidine, Cidex, 70% and 90% Isopropyl Alcohol, and 10%
(1:10) chlorine bleach to water solution.
CAUTION
• Do not immerse the connector on the cable in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
SPECIFICATIONS
Cable length
Patient module connector
Sensor connector
ENVIRONMENTAL
Storage/Transport Temperature
Operating Temperature
Storage/Transport Humidity
Operating Humidity
WARRANTY
Masimo warrants to the initial buyer only that this product, when used in accordance with the directions provided with the
Products by Masimo, will be free of defects in materials and workmanship for a period of six (6) months. Single use products
are warranted for single patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO TO BUYER.
MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL, EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION
ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. MASIMO'S SOLE OBLIGATION AND
BUYER'S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT MASIMO'S OPTION, TO REPAIR OR REPLACE
THE PRODUCT.
WARRANTY EXCLUSIONS
This warranty does not extend to any product that has been used in violation of the operating instructions supplied with the
product, or has been subject to misuse, neglect, accident or externally created damage. This warranty does not extend to any
product that has been connected to any unintended instrument or system, has been modified, or has been disassembled
or reassembled. This warranty does not extend to sensors or patient cables that have been reprocessed, reconditioned or
recycled.
IN NO EVENT SHALL MASIMO BE LIABLE TO BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT, SPECIAL OR
CONSEQUENTIAL DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED OF THE POSSIBILITY
THEREOF. IN NO EVENT SHALL MASIMO'S LIABILITY ARISING FROM ANY PRODUCTS SOLD TO BUYER (UNDER A CONTRACT,
WARRANTY, TORT OR OTHER CLAIM) EXCEED THE AMOUNT PAID BY BUYER FOR THE LOT OF PRODUCT(S) INVOLVED IN
SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY DAMAGES ASSOCIATED A PRODUCT THAT HAS BEEN
REPROCESSED, RECONDITIONED OR RECYCLED. THE LIMITATIONS IN THIS SECTION SHALL NOT BE DEEMED TO PRECLUDE
ANY LIABILITY THAT, UNDER APPLICABLE PRODUCTS LIABILITY LAW, CANNOT LEGALLY BE PRECLUDED BY CONTRACT.
NO IMPLIED LICENSE
Purchase or possession of this patient cable does not carry any expressed or implied license to use this cable with any device
that is not an authorized device or separately authorized to use SedLine cables.
3 feet (0.9 meters)
9 pin proprietary
Custom connector
-40° C to +70° C, ambient humidity
+5° C to +40° C, ambient humidity
10% to 95%, non-condensing
10% to 95%, non-condensing
4