Mensajes de error y eventos
5 Mensajes de error y eventos
Error
Pantalla oscura
Insufficient load
Overload
30
Causa
•
Back lighting set too dark
•
No mains voltage
•
Unit switched off
•
Mains cable not plugged in
•
Brief fault
•
Load plate not on the scale
•
Weighing range not reached
•
Weighing range exceeded
•
Result not yet stable
•
Function not permissible
•
Zeroing not possible with
overload or insufficient load
•
No calibration
•
Printout not yet ended
•
Switching the weighing unit
impermissible during dynamic
weighing
•
EAROM checksum error
Remedio
➜ Set back lighting ( b.LIGHt)
brighter
➜ Check mains
➜ Switch on unit
➜ Plug in mains plug
➜ Switch device off and back on
again
➜ Place load plate on the scale
➜ Set to zero
➜ Unload scale
➜ Reduce preload
➜ If necessary adjust vibration
adapter or weigh dynamically
➜ Unload scale and set to zero
➜ Unload scale
➜ Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
➜ Calibrate scale
➜ Call METTLER TOLEDO Service
➜ End printout
➜ Repeat required action
➜ End dynamic weighing
➜ Switch weighing unit
➜ Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
➜ Call METTLER TOLEDO Service
Manual de usuario 22011481A 03/05