Mensajes de error y eventos

5 Mensajes de error y eventos
Error
Pantalla oscura
Insufficient load
Overload
30
Causa
Back lighting set too dark
No mains voltage
Unit switched off
Mains cable not plugged in
Brief fault
Load plate not on the scale
Weighing range not reached
Weighing range exceeded
Result not yet stable
Function not permissible
Zeroing not possible with
overload or insufficient load
No calibration
Printout not yet ended
Switching the weighing unit
impermissible during dynamic
weighing
EAROM checksum error
Remedio
➜ Set back lighting ( b.LIGHt)
brighter
➜ Check mains
➜ Switch on unit
➜ Plug in mains plug
➜ Switch device off and back on
again
➜ Place load plate on the scale
➜ Set to zero
➜ Unload scale
➜ Reduce preload
➜ If necessary adjust vibration
adapter or weigh dynamically
➜ Unload scale and set to zero
➜ Unload scale
➜ Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
➜ Calibrate scale
➜ Call METTLER TOLEDO Service
➜ End printout
➜ Repeat required action
➜ End dynamic weighing
➜ Switch weighing unit
➜ Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
➜ Call METTLER TOLEDO Service