Generator
IV Pole Mounting
Bracket
Charger Module
Care should be taken in selecting a stable IV pole which should have a wide base and locking-style casters. An
IV pole with five casters, located in a circular pattern with a diameter of at least 23 inches (58 cm), such as the
I.V. League Ventilator Stat-Stand™ model 1059 (or equivalent) is recommended.
The Generator should be mounted such that its top surface is no more than 50 inches (127 cm) from the floor.
Consult your biomedical department in the event there are any questions regarding the stability of the
intended IV pole and mounting location. Mounting the Generator to an unstable IV pole could present a tip
over hazard to staff or patients.