8.
After eight minutes, if the target torque (1.1 Nm) has not
been reached, the software suggests carrying out a new
test.
8a.
Click on "Yes" before expiry of the allotted time to carry
out a new test. The software will automatically calculate
the correct hydration to reach the target consistency at
1.1Nm.
8b.
Click on "No" to continue the current test.
9.
The test will continue for 45 minutes (in the case of the
standard CHOPIN+ protocol) and will stop automatically.
It is possible to stop the test at any time while it is running by
clicking on [Stop test]:
D. Results
The test results are presented in three parts:
1 the graph of the test shown in green, along with the specific values (C and gradient)
2 a table summarizing the characteristic points of the curve
3 the profiler, a translation into six qualitative index of the graphical test result
Once the test has completed, it is possible to:
Save the test results
Print the test results
Delete the test results
Export the test results
Create a screen capture of the test results
1